64 09 713 D3610
32
D3610
.
201.01.08
.
02 09.2016
4 Installation
Sirona Dental Systems GmbH
4.5 Connect sensor
Operating Instructions and Installation XIOS XG USB module and sensors
4.5
Connect sensor
A calibration file is saved on the XIOS XG Select and XIOS XG Supreme
sensors which is transferred to the PC the first time that the sensor is
used.
✔
The USB module can be connected to the USB port of a PC either
directly or via a USB hub. The orange LED on the USB module lights
up, see also "Determine unit status" [ → 36].
1. Start SIDEXIS XG / SIDEXIS 4.
2. Plug the connector of the sensor into the front socket of the USB
module.
3. Register a patient in SIDEXIS XG / SIDEXIS 4 and establish
readiness for intraoral exposures.
A sensor calibration file is installed automatically if you are using
the XIOS XG Select or XIOS XG Supreme sensor on this PC for
the first time. If this does not occur and no exposure readiness is
established, disconnect the sensor from the USB module and
plug the cable in again.
4.6
Complete test exposures/acceptance tests
A test exposure must be completed after installing a XIOS XG USB
module or a new sensor.
An acceptance test must be completed by a service engineer from a
specialist retailer in Germany (Austria and Switzerland).
For completing test exposures, use a test phantom, see Consumables
and spare parts [ → 69]. Do not complete test exposures on a patient.