User manual for Digital Indicator and cooperating software
3.
TECHNICAL DATA
Power supply voltage
(depending on version)
External Fuse (required)
Power consumption
85...
230
...260V AC/DC; 50 ÷ 60 Hz
or 19...
24
...50V DC; 16V...
24
...35V AC
T - type, max. 2 A
typically 7 VA; max. 12 VA
Measurement inputs:
up to 8 inputs 0-20 mA or 4-20 mA, common ground
Sensor power supply output
24V DC ± 5% / max. 200 mA,
Not insulated from measurement inputs
Measurement range
± 9999 plus decimal point
Current measurement accuracy
± 0,25% ± one digit (for 0-20 mA range)
Accepted prolonged input overload: 20%
Measurement input resistance
100
Ω
Communication interface
Baud rate
RS 485, 8N1 / Modbus RTU,
not separated
1200 bit/s ÷ 115200 bit/s
Display
LCD graphic display, 128 x 64 points, with backlight
Protection level
IP 65 (from front, after using waterproof cover)
IP 40 (from front)
IP 20 (housing and connection clips)
Housing type
Housing material
Housing dimensions
Mounting hole
Assembly depth
Panel thickness
panel
NORYL - GFN2S E1
96 x 96 x 100 mm
90,5 x 90,5 mm
102 mm
max. 5 mm
Operating temperature
Storage temperature
Humidity
Altitude
0°C to +50°C
-10°C to +70°C
5 to 90% no condensation
up to 2000 meters above sea level
Screws tightening max. torque
0,5 Nm
Max. connection leads diameter
2,5 mm
2
Safety requirements
according to: PN-EN 61010-1
installation category: II
pollution degree: 2
voltage in relation to ground: 300V AC
insulation resistance
: >20M
Ω
insulation strength between power supply and
input/output terminal: 1min. @ 2300V (see Fig. 3.1)
EMC
PN-EN 61326
Software
(Windows environment)
•
, PC software for data logger easy configuring