User manual for Digital Indicator and cooperating software
or in the form of a bar (percentage) depending on the settings in the "
Channels list options
"
menu. Graphic indicators (bars) always show the measurement result to the
nominal
measurement range
ratio.
Fig. 6.15. Unit in "
Channels list
" results presentation mode.
"
Channels list options
" parameter set as "
bar graphs
".
Exceeding the
nominal measurement range
is signalled on the bars in the form of an arrow
indicating the direction of exceeding the range. If the measurement result for a given channel
exceeds the
allowable measurement range
, instead of the current result in numeric form the
display shows a "
-Hi-
" or "
-Lo-
" message (depending on the direction of exceeding, see
description of "
Upper ext.
" and "
Lower ext.
" parameters in "
Inputs settings
" menu). Press
the
[ESC/MENU]
key to go to "
Channels list options
" menu.
Fig. 6.16. Unit in "
Channels list
" results presentation mode.
"
Channels list options
" parameter set as "
values
".
Functions of keys in "Channels list" mode:
•
go to main menu (press and hold for approx. 2 seconds),
•
Go to "
Channels list options
" menu, description on page 27.
•
change channel number.
•
change results presentation mode.
7.
DEVICE PROGRAMMING
Defining the method of displaying measurement results in the selected results
presentation mode can be done in the presentation mode menu. The meaning of individual
parameters available for selected modes of results presentation is described in the
22
MENU
ESC
current time
channel number
channel name
percentage rate of measure result to
nominal input range
current date
measure result in graphical
mode (bargraph)
arrow informs about direction of
nominal range exceeding
warning message
informing that input allowable
range is exceeded
measure result in
numerical mode
unit name