To quit the menu without saving, press
CLEAR
.
Single ramp configuration?
The figure below illustrates the type of single ramp that can be generated and their parameters:
The
LOW level
and
HIGH level
levels are expressed:
•
As a percentage of the gauge if the scaling mode is ON.
•
In Volt or in temperature units if the scaling mode is OFF and according to the type of value emitted
(voltage or temperature emission).
The
Total Duration
corresponds to the amount of increase time required to go from the
Low level
to the
High
level
(and vice versa in the case of a decrease). It is expressed in seconds and the max time is limited to 1000s.
The
Time frame
corresponds to the amount of time you can have between pressing the emission start key and the
actual starting of generation. It is expressed in seconds and the max time is limited to 1000s.
Use the
F2
key to access the
CONFIGURATION/RAMP
menu. Attention: the appropriate function mode must be
programmed (
INCREASES
mode) to access the
CONFIGURATION/RAMP/SINGLE
menu.
In the
CONFIGURATION/RAMP/SINGLE
menu, use the
F2
key to go to the next field.
Use the navigation keys to enter the value:
•
As a percentage of the gauge if the scaling mode is ON.
•
↑
↑
↑
↑
and
↓
↓
↓
↓
to increase or decrease the value
•
and
to choose the hundreds/tens/units/decimal place/hundreds/thousands.
Press
VAL
to save the parameters.
To quit the menu without saving, press
CLEAR
.
Programming parameters in
Single Ramp
mode
T0
Tdt
T0
Td
Tdt
T0
Tdt
NH
NB
Td
To : time pressing the generation start key
Td :
Time frame
Tdt :
Total Time
NB :
Low Level
NH :
High Level