D.5.2
Configuration of the ramp generation
The CONFIGURATION/RAMP menu is used for the generation of ramps by increases, single or cyclic.
Ramp by increases configuration?
The figure below illustrates the type of single ramp that can be generated and their parameters:
The
LOW level
and
HIGH level
levels are expressed:
•
As a percentage of the gauge if the scaling mode is ON.
•
In mVolt or in temperature units if the scaling mode is OFF and according to the type of value emitted
(voltage or temperature emission).
The
Duration
corresponds to the amount of increase time required to go from the
Low level
to the
High level
(and
vice versa in the case of a decrease). It is expressed in seconds and the max time is limited to 1000s.
The
Time frame
corresponds to the amount of time you can have between pressing the emission start key and the
actual starting of generation. It is expressed in seconds and the max time is limited to 1000s.
Use the
F2
key to move to the next field.
Use the navigation keys to enter the value:
•
As a percentage of the gauge if the scaling mode is ON.
•
↑
↑
↑
↑
and
↓
↓
↓
↓
to increase or decrease the value
•
and
to choose the hundreds/tens/units/decimal place/hundreds/thousands.
Press
VAL
to save the parameters.
Programming parameters in
Increases
mode
T0
Tdt
T0
Td
Tdt
T0
Tdt
NH
NB
Td
INC
To : time pressing the generation start key
Td :
Time frame
Tdt :
Duration
NB :
Low Level
NH :
High Level
INC :
Increase
(value of increase as a %, in mVolt, °C or °F)