B.1.6
Operating modes
There are 2 different operating modes:
•
Thermocouple measurement (displayed in mV or °C or °F),
•
Simulation of thermocouple/voltage emission (value displayed in mV or °C or °F).
The functional and electrical characteristics not to be exceeded are described below:
B.1.6.1
Constant voltage measurement
Range
+100mV
Resolution (display)
1 µV or 0.01 °C or 0.01 °F
Scope of range
-10 mV to + 100 mV
Scaling
yes
B.1.6.2
Voltage emission
Range
+ 80 mV
Resolution (display)
1 µV or 0.01 °C or 0.01 °F
Scope of range
-9.5 mV to +80 mV
Scaling
yes
B.1.6.3
Electrical characteristics not to be exceeded.
Function
Range
max V
in
Z
load
U
measurement
100 mV
60 V
U
emission
80 mV
1000
Ω
Ω
Ω
Ω
min