![SIKA UC TC Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/sika/uc-tc/uc-tc_manual-instruction_1270468021.webp)
SIKA, Dr. Siebert & Kühn GmbH & Co. KG, Struthweg 7-9, 34260 Kaufungen, Germany
21/40
Tel: +49-5605-803-0, Fax: +49-5605-803-54, email: [email protected], web: www.SIKA.net
The
Define/parameters
menu makes it possible to define the format (Number of digits displayed) and unit.
Once the parameters have been set, the scaling is automatically enabled. To disable it, enter the
Configuration/Scaling
menu, select
OFF
and confirm by pressing the
VAL
key.
D.2
Nulling
The relative measurement function available on the device makes it possible to cancel a constant or spurious value
via programming.
When the relative measurement function is enabled, the symbol
is displayed on the measurement screen.
The
NULL/define
menu makes it possible to program the value of the Tare (positive or negative). This value is
obtained from the measurements:
Value Displayed = Value measured – Value of the Tare
Scaling
pictogram
Pictogram of the
TARE mode