FMP1
Basic setting and Operating
Technical changes reserved
- 49 -
Digital input
:
The digital input parameter can be set to one of three available functions.
Set the value for the desired function.
Available values: "1" = ZERO function, "2" = PEAK function, "3" = HOLD function
Connect a switch or button to the digital input, observing the instructions for electrical
connection (
§ 5.5).
Sensor input sensitivity (force sensors or load cells):
This parameter allows the FMP1 to be adapted to force sensors or load cells with output sig-
nals from 1000 to 3000 mV.
Set the sensitivity parameter to and confirm the entry.
Set the desired full-scale value.
Value range: 1000…3000 mV.
6.5
Device calibration
Use the following procedure to adapt the device to special sensors or requirements:
Set the full scale value "FS" of the device to the nominal load of the sensors being used.
e.g.: 1 load cell rated 1000 kg
--->
"FS" = 1000
2 load cells each rated 1000 kg --->
"FS" = 2000.
Set the sensitivity "SE" of the analogue input.
e.g.: Sensor output 1 mV
--->
"SE" = 1.000
Sensor output 1.5 mV
--->
"SE" = 1.500.
Using the tare parameter or the ZERO function, set the display reading to zero.