Analogue input
:
You can adjust the analogue input to match the characteristics of the connected transducer.
Set the value for the characteristic of the analogue input.
Available values: "1" = 2 mV/V, "2" = 3 mV/V, "3" = 4…20 mA or 0…10 V (on request).
Analogue output
:
Configuration of the analogue output starts with this parameter. First set the desired output
signal. Then you can adjust the other parameters – zero offset, full scale and amplitude – of
the analogue output to match your requirements.
Set the parameter to and confirm the entry.
Set the value for the desired output signal.
Available values:
"diS" = disabled, "Pn10" =
10 V, "0-20" = 0…20 mA , "4-20" = 4…20 mA .
Zero offset:
Use this parameter to compensate for the zero offset of the analogue output. It can be up to
30 to 50% of the full-scale value of the device.
First reset the device by pressing the ZERO key (in measuring mode) or setting the tare
parameter as described above.
Set the parameter to and confirm the entry.
Set the desired zero offset value.
Value range: 0000…9999.
Full-scale value of analogue output:
Use this parameter to change the full-scale value of the analogue output. The change is pro-
portional to the full scale value of the FMP1.
Set the parameter to and confirm the entry.
Set the desired full-scale value.
Value range: 0000…9999.
Analogue output amplitude:
Use this parameter to adjust the amplitude of the analogue output. The adjustment range is 5
to 11.5 V with voltage output or 12 to 28 mA with current output.
Load the device sufficiently to attain the desired full scale value of the analogue output.
Set the parameter to and confirm the entry.
Set the desired amplitude value.
Value range: 000.0…109.0 %.
Notice: For more precise adjustment, the final digit changes only after 10 keypresses.
Notices:
After device switch-on, the analogue output is activated only after the readout has
stabilised.
The response time of the analogue output is affected by the digital filter setting. We
recommend using a filter setting
≥
2 for the analogue input..
Содержание FMP1
Страница 26: ...Technische Daten FMP1 26 SIKA Ba_FMP1 03 2014 10 1Abmessungen Aluminiumgeh use Frontplattenausschnitt...
Страница 27: ...FMP1 Technische nderungen vorbehalten 27 F r ihre Notizen...
Страница 54: ...Technical Data FMP1 54 SIKA Ba_FMP1 03 2014 10 1Dimensions Aluminium housing Front panel cut out...
Страница 55: ...FMP1 Technical changes reserved 55 For your notes...