FMP1 Electrical
connection
Technical changes reserved
- 39 -
5.2
Force sensors or load cells
Sensors with strain gauge signals are connected to the 14-way terminal strip. Several sen-
sors can be wired in parallel – up to 4 x 350
Ω
or 8 x 700
Ω
, depending on the sensor imped-
ance. Use 4-wire connection.
Terminal assignment:
Terminal
Designation
Colour *
1 (S+)
Strain gauge output signal (+)
white
2 (S-)
Strain gauge output signal (-)
yellow
3 (-AL)
Strain gauge supply voltage (-)
black
4 (+AL)
Strain gauge supply voltage (+)
red
The shield is connected to terminal 3.
* The colours refer to SIKA sensors with standard cables.
5.3
Relay outputs
The floating relay contacts are available on terminals 10 to 14 of the 14-way terminal strip.
The relays are controlled by the programmable switching points SP1 and SP2.
CAUTION! Maximum contact rating: 115 V
AC
, 0.2 A or 48 V
DC
, 0.2 A!
The maximum contact rating is only valid for purely resistive loads and must never be
exceeded. Switching inductive or capacitive loads in particular may cause large voltage
or current spikes.
Take suitable protective measures for your application.
Terminal assignment:
Terminal
Relay
Contact type
10 1
Normally
closed
11 1
Normally
open
12 2
Normally
closed
13 2
Normally
open
14
1 and 2 Common terminal
Содержание FMP1
Страница 26: ...Technische Daten FMP1 26 SIKA Ba_FMP1 03 2014 10 1Abmessungen Aluminiumgeh use Frontplattenausschnitt...
Страница 27: ...FMP1 Technische nderungen vorbehalten 27 F r ihre Notizen...
Страница 54: ...Technical Data FMP1 54 SIKA Ba_FMP1 03 2014 10 1Dimensions Aluminium housing Front panel cut out...
Страница 55: ...FMP1 Technical changes reserved 55 For your notes...