Instruction Manual PhaseGuard C/ T/ HT
Initial start-up
18
11026E/4
WORKSTEP
PROCEDURE
Language:
0: German,
1: English – setting the
operating language
Current source channel:
0: Turb, 1: Grad,
Default = 0
Setting the source for the
current output
Current from:
0.000
Setting the lower value of
the current range
Current to:
100
Setting the upper value of
the current range
Limit source channel:
0: Turb, 1: Grad,
Default = 0
Setting the source for limit
formation
Limits mode:
0: inactive;
1: exceeded; 2: undershot;
default = 0
Upper limits:
0.000
Enter limits with upper
threshold.
Lower limits
: 0.000
Enter limits with lower
threshold.
Integration:
0 s
Integration time of the
measuring value of
0 .. 60000 seconds.
Output 1 invert:
0: No, 1: Yes
Invert: inverts output 1
(alarm)
Output 2 invert:
0: No, 1: Yes
Invert: inverts output 2
(limit)
Factory set.:
0: No; 1: Yes
With
Yes
the factory set-
ting is loaded.
4.3: Close the file after the entry. Changes are saved au-
tomatically.
The window of the removable disk disappears and then
reappears in a moment.
4.4: If the configuration is successful, the file
Config.OK
appears.
If the configuration is not successful, the file
Config.ERR
appears.
The configuration may fail if the entered values
are not within the permitted range or if text areas have
been deleted or changed.
If the configuration is faulty, the file
Config.txt
is auto-
matically generated again and the procedure can be re-
peated.
5.
Carry out zero calibration
as described in Section 6.2.
6.
Complete procedure.
After successfully setting all parameters, remove the USB
connector from the instrument and screw on the cover
again.
If malfunctions occur, please refer to the Reference Manual.