Mechanical mounting
Instruction Manual PhaseGuard C/ T/ HT
11026E/4
13
4.
Mechanical mounting
The instrument can be installed in horizontal or vertical sample lines using the standards-
compliant in-line housing.
The mounting dimensional drawing (PHASEGUARD-MB) provides detailed information about
the dimensions.
Figure 3: Position of the sensor in horizontal mounting
Photometer
Centering pin on the instrument
Sensor with measurement opening
pointing downward
Varivent® sample line
The photometer must be mounted in the line at least 0.5 m away from sight glasses
and other unwanted light sources.
In the vertical mounting position, the centering pin (2, Figure 3) must point upward
(X, Figure 4).
In the horizontal mounting position, the centering pin (2, Figure 3) must point to the
right (Y, Figure 4).
Basics of mounting
the photometer