Instruction Manual PhaseGuard C/ T/ HT
Instrument description
8
11026E/4
2.2.
Technical data
TURBIDITY MEASUREMENT
VALUES
Measuring principle
Absorption
at 880 nm (T/HT)
at 430 nm (C/HC)
Measurement span
0 .. 100% absorption
Measuring range
0 .. 100% absorption
Sample temperature
-10 .. +100°C
120°C max. 2 hr
Ambient temperature
-10 .. +50°C
Pressure
1 MPa (10 bar) / 100°C
Resolution
0.5% absorption
Warm-up time
< 3 min
Repeatability (2 measurements
with 1 instrument)
0.5% absorption
Linearity
0.5% absorption
Temperature stability
±0.05%/K (full scale)
Reaction time
< 0.3 s at 0 seconds integration time (step response
limit switch)
Ambient humidity
0 .. 100% relative humidity
Path lengths
10 mm (T/C)
5 mm (HC/HT)
Outputs
1 x 4 .. 20 mA, galvanically separated up to max.
35 VDC relative to ground and max. 500 Ω burden
2 x switching transistor with open collector
U
EXT
3 .. 30 VDC, I
L
max. 8 mA
Interfaces
USB (for configuration via text file)
Modbus RTU / Profibus DP (both optional)
To optional SICON control unit