background image

 

Document number: 11026E 

Version: 4 

Valid from: Software Vx26 

INSTRUCTION MANUAL

 

PhaseGuard C/ T  /HT 

 

In-line Interface Monitor 

 

 

Содержание PhaseGuard C

Страница 1: ...Document number 11026E Version 4 Valid from Software Vx26 INSTRUCTION MANUAL PhaseGuard C T HT In line Interface Monitor...

Страница 2: ...IGRIST PHOTOMETER AG subject to changes without notice 9 2016 SIGRIST PHOTOMETER AG Hofurlistrasse 1 CH 6373 Ennetb rgen Switzerland Tel 41 41 624 54 54 Fax 41 41 624 54 55 info photometer com www pho...

Страница 3: ...trical installation 14 5 2 Installation of the PhaseGuard 15 5 3 Installation of the SICON control unit field bus 16 6 Initial start up 17 6 1 Commissioning procedure 17 6 2 Perform zero calibration 1...

Страница 4: ...Instruction Manual PhaseGuard C T HT ii 11026E 4 Terms used in this document glossary Please refer to our website for specialist terms www photometer com...

Страница 5: ...first time registration re quired 11018E Data sheet Descriptions and technical data about the pho tometer 11052E Service Manual Repair and conversion instructions for service engineers 11050DEF Decla...

Страница 6: ...e may cause explosions resulting in serious material damage and death Warning about bodily injury or hazards to health with long term effects Non observance of this warning may lead to injuries with p...

Страница 7: ...ontrol system Customer side control system Varivent in line housing or compatible optional Sample line 2 1 2 Proper use The PhaseGuard and its peripherals are designed to detect phase transitions in l...

Страница 8: ...ggressive media may cause leakage that could be lethal to persons in the vicinity It is not permitted to use the measuring instrument for applications in which the medium can attack it Parts which con...

Страница 9: ...ation of the PhaseGuard The photometer displays the following rating plate Figure 2 PhaseGuard rating plate Manufacturer Country of origin Product name Serial number Date of manufacture Service voltag...

Страница 10: ...with mid range turbidities 1 119788 PhaseGuard T Sensor head made of Hastelloy 1 118678 PhaseGuard HT Length of sensor head 56 5 mm Path length 5 mm Detection of phase tran sitions with high turbidi t...

Страница 11: ...ger housing the two cable pene trations and M12 connector Profibus DP 118696 RTU Modbus 1 118342 SICON control unit 1 119040 SICON M Multichannel control unit 1 120444 Device cable 8 pole 10 m with co...

Страница 12: ...up time 3 min Repeatability 2 measurements with 1 instrument 0 5 absorption Linearity 0 5 absorption Temperature stability 0 05 K full scale Reaction time 0 3 s at 0 seconds integration time step resp...

Страница 13: ...w Sapphire Weight Approx 2 1 kg Approx 2 5 kg PhaseGuard with interface card Dimensions 120 x 143 mm for instrument types T and C 120 x 164 5 mm with interface card 120 x 166 5 mm for instrument type...

Страница 14: ...s is installed customer side Before commissioning clarify the following points Position of the on off switch and the way it works Where and how the sample supply can be stopped Who the responsible aut...

Страница 15: ...o incorrect power supply If the instrument is connected to an incorrect service voltage it can be damaged The instrument may be connected only to voltage sources as specified on the rating plate Damag...

Страница 16: ...similar items are available and are put to use To avoid faulty measurement values as much as possible the maintenance duties must be carried out according to the servicing schedule Evaluate digital ou...

Страница 17: ...ns Figure 3 Position of the sensor in horizontal mounting Photometer Centering pin on the instrument Sensor with measurement opening pointing downward Varivent sample line The photometer must be mount...

Страница 18: ...cal installations must be observed at all times Further the following basic principles must be observed The power supply 9 30 VDC is to be provided by the customer It must be stable low noise voltage...

Страница 19: ...color for device cable art no 120444 Remarks GND supply 1 white Supply 9 30 VDC 2 brown RS 485 A 7 blue Serial interface for SICON M Alternative initiate zero calibration Reference Manual RS 485 B 5...

Страница 20: ...ds on the power supply and the cross section in use CROSS SECTION MAX DISTANCE FOR 12 VDC WITH PROFIBUS MAX DISTANCE FOR 24 VDC WITH PROFIBUS REMARKS mm2 m m 0 14 40 20 150 100 0 20 60 30 250 150 Stan...

Страница 21: ...3 2 Connect the PhaseGuard via USB cable to the PC The PhaseGuard registers itself as a removable disk in the Windows operating system 3 3 Depending on the Windows configuration a win dow opens autom...

Страница 22: ...ts output 1 alarm Output 2 invert 0 No 1 Yes Invert inverts output 2 limit Factory set 0 No 1 Yes With Yes the factory set ting is loaded 4 3 Close the file after the entry Changes are saved au tomati...

Страница 23: ...without previously emptying the sample line The photometer can be removed from a line only when the line is completely empty If this is not the case flooding and damage to equipment as well as bodily...

Страница 24: ...iguration is not successful the file Zero ERR appears In this case repeat adjustment and check the points in the following list one after the other Soiled window on the sensor head Reference Manual Ai...

Страница 25: ...Change desiccant Section 8 2 Obligatory measure for maintaining the measuring accuracy and for protecting the electronics Interval depends on op erating and ambient conditions Annually or as needed Op...

Страница 26: ...least at room temperature before it is opened If the desiccant has to be frequently replaced the sealing of the instrument should be checked by a SIGRIST service technician The following servicing dut...

Страница 27: ...TEP PROCEDURE 1 Empty sample line Make sure the line is empty 2 Put the system into a safe state and remove the in strument from the line Danger when removing the photometer if the sample line has not...

Страница 28: ...age to the sensor head due to improper handling No other cleaning methods or other cleaning agents should be used Clean the two windows on the sensor head X with a mild non abrasive cleaning agent e g...

Страница 29: ...y may occur WORKSTEP PROCEDURE 1 Empty sample line Ensure that the sample line is empty 2 Put the system into a safe state and remove the in strument from the line Danger when removing the photometer...

Страница 30: ...he sample line 3 1 Remove lock clamp for fastening the blanking plate 3 2 Remove blanking plate and replace the old gasket arrow with a new one 4 Re mount photometer and blanking plate Re mount the ph...

Страница 31: ...head Reference Manual No display or indication only with SICON Check whether the service voltage is correctly connect ed to the SICON Reference Manual If the listed measures do not result in the desi...

Страница 32: ...e the instrument Connecting electrical lines is extremely dangerous Parts of the system can also be damaged Local regulations for electrical installations must be observed at all times Interrupt the p...

Страница 33: ...rt The original packaging materials of the photometer should be used for packaging the pho tometer and its peripheral components if possible If the original packaging is no longer available note the f...

Страница 34: ...able CATEGORY MATERIALS DISPOSAL POSSIBILITIES Packaging Cardboard paper Reuse as packaging material local dis posal points incineration plants Protective foils polystyrene shells Reuse as packaging m...

Страница 35: ...NO ARTICLE NAME REMARKS 108247 O ring EPDM 60 x 3 75 shore A Material EPDM 60 x 3 mm 111391 Desiccant bag 30 g Sealed packaging unlimited shelf life 112379 O ring NBR 60 x 3 70 shore A Material Nitril...

Страница 36: ...Appendix Instruction Manual PhaseGuard C T HT 32 11026E 4 15 Appendix 15 1 Dimension sheet PhaseGuard MB...

Страница 37: ...Appendix Instruction Manual PhaseGuard C T HT 11026E 4 33 15 2 Overview of gaskets on the PhaseGuard...

Страница 38: ...Appendix Instruction Manual PhaseGuard C T HT 34 11026E 4 15 3 PhaseGuard connection diagram...

Страница 39: ...ropean Union 4 Explanation of symbols 2 G Guidelines 4 H Horizontal 13 I In line housing 13 Internet 27 L Limits definition 18 M Mounting 14 Mounting position 14 P Packaging 29 Pictograms 2 Proper use...

Страница 40: ...teuer module Adressbl ock 0 mod_1341480540857_6 docx 4061 1 SIGRIST PHOTOMETER AG Hofurlistrasse 1 CH 6373 Ennetb rgen Switzerland Tel 41 41 624 54 54 Fax 41 41 624 54 55 info photometer com www photo...