SIGLENT Remote Manual
89 SDM3045X
2.10 SYSTem Subsystem
2.10.1 SYSTem:BEEPer:STATe {ON|1|OFF|0}
SYSTem:BEEPer:STATe?
Disables or enables the beep heard during continuity, diode, or Probe Hold measurements,
or when an error is generated from the front panel or remote interface.
Parameter
Typical Return
{ON|1|OFF|0}
,
default ON
0(OFF)or1(ON)
Disable keyboard click:
SYST:BEEP:STAT OFF
This command affect the beeper heard associated with diode and continuity tests,
errors and so on.
This parameter is set to its default value after a Factory Reset.
2.10.2 SYSTem:COMMunicate:LAN:IPADdress "<address>"
SYSTem:COMMunicate:LAN:IPADdress? [{CURRent|STATic}]
Assigns a static Internet Protocol (IP) address for the instrument.
Parameter
Typical Return
Command
:
"nnn.nnn.nnn.nnn"
Query
:
{CURRent|STATic}
,
default CURRent
Default
:
169.254.3.5
“192.168.1.101”
Set a static IP address:
SYST:COMM:LAN:IPAD "10.11.13.212"
This setting is non-volatile; it is not changed after a Factory Reset.