2.9.36 [SENSe:]CAPacitance:RANGe {<range>|MIN|MAX|DEF}
[SENSe:]CAPacitance:RANGe? [{MIN|MAX|DEF}]
Manually choose measurement range.
Parameter
Typical Return
<range>
:
{2nF|20nF|200nF|2uF|20uF|200uF|10000uF|AUTO}
Default 2uF.
+2.00000000E-06
Using the 2uFfileconfigurationcapacitance to measure. Make and read two measurements.
CONF:CAP
CAP:RANG 2E-6
SAMP:COUN 2
READ?
Typical Response
:
+7.28283777E-10,+7.28268544E-10
Choose the fixed range ([SENSe:]<function>:RANGe) ,disable the automatic
adjustment range.
If the input signal is greater than can be measured on the specified manual range, the
instrument displays the word Overload on front panel and returns "9.9E37" from the
remote interface.
This parameter is set to its default value after a Factory Reset or CONFigure function.