![SIGLENT TECHNOLOGIES CO.,LTD SDM3045X Скачать руководство пользователя страница 50](http://html1.mh-extra.com/html/siglent-technologies-co-ltd/sdm3045x/sdm3045x_remote-manual_1267548050.webp)
Remote Manual
SIGLENT
SDM3045X 48
2.8.2 MEASure:CURRent:{AC|DC}? [{<range>|AUTO|MIN|MAX|DEF}]
Sets all measurement parameters and trigger parameters to their default values for AC or
DC current measurements and immediately triggers a measurement. Also specifies the
stalls through the incoming parameters.
Parameter
Typical Return
<range>
:
{600uA|6mA|60mA|600mA|6A|10A|AUTO}
,
default AUTO(autorange).The 600uA and 6mA file can only
be set in DC current mode,AC mode minimum gear for
60mA
None
Configure AC current measurement using the 6A range.Make and read two
measurements:
CONF:CURR:AC 6
SAMP:COUN 2
READ?
Typical Response
:
+4.32133675E-04,+4.18424606E-04
You can let autoranging select the measurement range, or you can manually select a
fixed range. Autoranging conveniently selects the range for each measurement
based on the input signal. For fastest measurements, use manual ranging
(autoranging may require additional time for range selection).
Autoranging (AUTO or DEFault), will generate an error if you specify a <
resolution
>
because the instrument cannot accurately resolve the integration time (especially if
the input continuously changes). If your application requires autoranging, specify
DEFault for the <
resolution
> or omit the <
resolution
> altogether.
If the input signal is greater than can be measured on the specified manual range, the
instrument displays the word Overload on front panel and returns "9.9E37" from the
remote interface.