Remote Manual
SIGLENT
SDM3045X 10
1.5 R? [<max_readings>]
Reads and erases all measurements from the reading memory up to the specified
<
max_readings
>. The measurements are read and erased from the reading memory
starting with the oldest measurement first.
Parameter
Typical Return
1 to 10,000
readings
Default is all
readings in
memory
#247-1.06469770E-03,-1.08160033E-03,-1.22469433E-03
The "#2" means that the next 2 digits indicate how many characters will
be in the returned memory string. These two digits are the "47" after the
"#2". Therefore, the remainder of the string is 47 digits long:
-1.06469770E-03,-1.08160033E-03,-1.22469433E-03
Read and remove the three oldest readings:
TRIG:COUN 3
INIT
R? 3
The R? and DATA:REMove? queries allow you to periodically remove measurements
from the reading memory that would normally cause the reading memory to overflow.
You can store up to 1,000 measurements in the reading memory of the SDM3055x. If
reading memory overflows, new measurements will overwrite the oldest
measurements stored; the most recent measurements are always preserved. No
error is generated, but the Reading Mem Ovfl bit (bit 14) is set in the Questionable
Data Register's condition register (see Status System Introduction).
The instrument clears all measurements from reading memory when the
measurement configuration changes, or when any of these commands is executed:
INITiate
MEASure:<function>?
READ?