SIGLENT Remote Manual
35 SDM3045X
2.6 CONFigure Subsystem
The CONFigure commands are the most concise way to configure measurements.These
commands use default measurement configuration values. However, these commands do
not automatically start measurements, so you can modify measurement attributes before
initiating the measurement.
Use INITiate or READ? to initiate measurements.
Command Summary
CONFigure?
CONFigure:CONTinuity
CONFigure:CURRent:{AC|DC}
CONFigure:DIODe
CONFigure:{FREQuency|PERiod}
CONFigure:{RESistance|FRESistance}
CONFigure:TEMPerature
CONFigure[:VOLTage]:{AC|DC}
CONFigure:CAPacitance
Default Settings for the CONFigure Command
The CONFigure commands select the function, range and sampling rate in one command.
All other parameters are set to their default values (below).If no range and sampling
speed settings, which will restore the default value.
Measurement Parameter
Default Setting
Autozero
OFF
Range
AUTO
Samples per Trigger
1 sample
Trigger Count
1 trigger
Trigger Delay
1 trigger
Trigger Source
Immediate
Trigger Slope
NEGative
Math Functions
Disabled
Using CONFigure
The following example uses CONFigure and READ? to make an externally-triggered
measurement. The CONFigure command configures DC voltage measurements but does
not place the instrument in the "wait-for-trigger" state.
The READ? query places the instrument in the "wait-for-trigger" state, initiates a
measurement when the rear-panel
Ext Trig
input is pulsed (low by default), stores the
measurement in reading memory, and transfers the measurement to the instrument's
output buffer. The default range (auto range) and resolution (0.3 PLC) are used for the
measurement.