Remote Manual
SIGLENT
SDM3045X 36
CONF:VOLT:DC
TRIG:SOUR EXT
READ?
Typical Response: -5.21391630E-04
The following example is like the previous one, but it uses INITiate and FETCh? instead of
READ?. The INITiate command places the instrument in the "wait-for-trigger" state,
triggers a measurement when the rear-panel
Ext Trig
input is pulsed (low by default), and
sends the measurement to reading memory. The FETCh? query transfers the
measurement from reading memory to the instrument's output buffer.
CONF:VOLT:DC
TRIG:SOUR EXT
READ?
Typical Response: -5.21205366E-04
Storing measurements in reading memory with INITiate is faster than sending
measurements to the instrument's output buffer using READ? (provided you do not send
FETCh? until done). The INITiate command is also an "overlapped" command. This
means that after executing INITiate, you can send other commands that do not affect the
measurements. This allows you to check for data availability before initiating a read
attempt that might otherwise time out. Note that the FETCh? query waits until all
measurements are complete to terminate. You can store up to 1,000 measurements in the
reading memory of the SDM3055.
The following example configures the instrument for 2-wire resistance measurements,
triggers the instrument to make one measurement using INITiate, and stores the
measurement in reading memory. The 10 kΩ range is selected.
CONF:RES 10000
INIT
FETC?
Typical Re5.2104