Remote Manual
SIGLENT
SDM3045X 60
[SENSe:]{FREQuency|PERiod}
This subsystem configures frequency and period measurements.
Command Summary
[SENSe:]{FREQuency|PERiod}:NULL[:STATe] {ON|1|OFF|0}
[SENSe:]{FREQuency|PERiod}:NULL:VALue {<value>| minimum | maximum | default}
[SENSe:]{FREQuency|PERiod}:NULL:VALue:AUTO {ON|1|OFF|0}
[SENSe:]{FREQuency|PERiod}:VOLTage:RANGe {<range>|MIN|MAX|DEF}
[SENSe:]{FREQuency|PERiod}:VOLTage:RANGe:AUTO {OFF|ON|ONCE}
2.9.8 [SENSe:]{FREQuency|PERiod}:NULL[:STATe] {ON|1|OFF|0}
[SENSe:]{FREQuency|PERiod}:NULL[:STATe]?
Enables or disables the null function for frequency and period measurements.
This parameter is shared between frequency and period measurements.
Parameter
Typical Return
{ON|1|OFF|0}
,
default OFF
0 (OFF) or 1 (ON)
Configure frequency measurements, using the null function to subtract 1kHz
from the measurements. Make and read two measurements:
CONF:FREQ
FREQ:NULL:STAT ON;VAL 1 kHz
SAMP:COUN 2
READ?
Typical Response
:
+1.1502,+1.0702
Enabling the scaling function also enables automatic null value selection ([SENSe:]
{FREQuency|PERiod}:NULL:VALue:AUTO ON).
To set a fixed null value, use this command:
[SENSe:]{FREQuency|PERiod}:NULL:VALue.
The instrument disables the null function after a Factory Reset or CONFigure
function.