SIGLENT Remote Manual
29 SDM3045X
2.4.4 CALCulate:SCALe:REFerence:AUTO {ON|1|OFF|0}
CALCulate:SCALe:REFerence:AUTO?
Enables or disables automatic reference selection for the dB scaling functions
Note:This command applies only to ACV and DCV measurement functions.
Parameter
Typical Return
{ON|1|OFF|0}, default ON
0(OFF)or1(ON)
Enable the DB function with automatic reference
selection and use the first measurement as the reference
value:
CALC:SCAL:DBM:REF 50
CALC:SCAL:FUNC DB
CALC:SCAL:REF:AUTO ON
CALC:SCAL:STAT ON
READ?
ON:
the first measurement made will be used as the reference for all subsequent
measurements, and automatic reference selection will be disabled:
For the dB scaling function, the first measurement is converted to dBm, and
CALCulate:SCALe:DB:REFerence is set to the result.
OFF:
CALCulate:SCALe:DB:REFerence specifies the reference for DB scaling.
The instrument enables automatic reference selection when the scaling function is
enabled (CALCulate:SCALe:STATe ON).
This parameter is set to its default value after a Factory Reset.