[SENSe:]TEMPerature Subsystem
This subsystem configures temperature measurements.
Command Summary
[SENSe:]TEMPerature:NULL[:STATe]
[SENSe:]TEMPerature:NULL:VALue
[SENSe:]TEMPerature:NULL:VALue:AUTO
[SENSe:]TEMPerature:TRANsducer?
[SENSe:]TEMPerature:{UDEFine|MDEFine}:{THER|RTD}:TRANsducer:LIST?
[SENSe:]TEMPerature:{UDEFine|MDEFine}:{THER|RTD}:TRANsducer
[SENSe:]TEMPerature:{UDEFine|MDEFine}:{THER|RTD}:TRANsducer:POINt?
2.9.19 [SENSe:]TEMPerature:NULL[:STATe] {ON|1|OFF|0}
[SENSe:]TEMPerature:NULL[:STATe]?
Enables or disables the null function for temperature measurements.
Parameter
Typical Return
{ON|1|OFF|0}
,
default ON
0(OFF)or1(ON)
Use the null function to subtract 25°.
TEMP:NULL:STAT ON;VAL 25
Enabling the scaling function also enables automatic null value selection
([SENSe:]TEMPerature:NULL:VALue:AUTO ON).
To set a fixed null value, use this command: [SENSe:]TEMPerature:NULL:VALue.
The instrument disables the null function after a Factory Reset or CONFigure
function.