Acquisition and interrupt handling software is loaded to the RAM
before execution, in order to attain maximum performance.
2.15 Reference Signal Generator
The Reference Signal Generator (RSG) consists of the following
blocks:
•
A Programmable Sine Wave Generator
•
A Programmable Gain Amplifier
Safety
Switch
Figure 2-6: Reference Signal Generator Block Diagram
The CPU controls the RSG in order to produce the required frequency
and amplitude during an acquisition cycle.
2.16 Probe Selection Board (PSB) Overview
The purpose of the PSB is to enable the operator to work
simultaneously with up to three probes. Currently the system is
supplied with two probes, large and small. The probes are connected
all the time, and by pressing the “GO” button on one of them the system
selects the probe, and connects its data lines to the PM board.