
Neck
ThinSliceSpi
2
nd
reconstr.
kV
120
Effective mAs
150
Slice collimation
0.75 mm
Slice width
5.0 mm
1.0 mm
Feed/Rotation
12.0 mm
Rotation time
0.75 sec.
Kernel
B30s
B30s
Increment
5.0 mm
0.7 mm
Image order
cr-ca
CTDI
w
12.6 mGy
Effective dose
Male: 1.7 mSv
Female: 1.9 mSv
93
This protocol can also be used in case of tumor suspi-
cious of the hypopharynx and larynx. It is recommended
to acquire the images in E-phonation, and reconstruct
the images with 1 mm slice width and increment of
0.7 mm for performing MPRs. This would help differ-
entiate tumor lesions from normal pharyngeal mucosa.
MPRthick: ThinSliceSpi
sagittal
Image thickness
3
Distance between images
3
Number of images
20
The 2
nd
reconstruction can be loaded in the 3D Card,
MPR. Select the Range function and a predefined MPR
range, linked to the series, will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
Contrast medium IV injection
Start delay
45 sec.
Flow rate
3 ml/sec.
Total amount
100 ml
Содержание SOMATOM Sensation 16
Страница 1: ...SOMATOM Sensation 16 Application Guide Routine Protocols Software Version A50 ...
Страница 7: ...7 Content ...
Страница 23: ...23 General ...
Страница 29: ...29 Children ...
Страница 49: ...49 Children ...
Страница 55: ...55 Children ...
Страница 94: ...94 Shoulder Overview ShoulderSpi Spiral mode for HiRes bone studies and soft tissues ...
Страница 115: ...115 Abdomen ...