AngioFast
2
nd
reconstr.
kV
120
Effective mAs
130
Slice collimation
1.5 mm
Slice width
5.0 mm
2.0 mm
Feed/Rotation
21.0 mm
Rotation time
0.5 sec.
Kernel
B30f
B20f
Increment
5.0 mm
1.5 mm
Image order
cr-ca
CTDI
w
9.1 mGy
Effective dose
Male: 10.8 mSv
Female: 16.4 mSv
Do not administer oral contrast medium, as this
impairs the editing of MIP/SSD/VRT images.
* CARE Bolus may be used to optimize the bolus
timing. Set the ROI for monitoring scan in the abdo-
minal aorta at the beginning of the scan range
with triggering threshold of 120 HU, or use manual
triggering.
MIPthin: AngioFast
sagittal
Image thickness
3
Distance between images
3
Number of images
60
For the 2
nd
reconstruction the Autoload into the
MIPthin Range on the 3D Card is activated. The images
will be automatically loaded into 3D, MIPthin, and
sagittal MIPthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
Contrast medium IV injection
Start delay
20 sec.*
Flow rate
3.0 ml/sec.
Total amount
120 ml
123
Abdomen