30
Children
HeadSpi
Indications:
Spiral mode for routine head studies, e.g. tumors,
hydrocephalus, hemorrhaging, abnormalities, etc.
HeadSpi
kV
120
Effective mAs
*
Slice collimation
1.5 mm
Slice width
4.0 mm
Feed/Rotation
12.2 mm
Rotation time
0.75 sec.
Kernel
C30s
Increment
4.0 mm
Image order
ca-cr
Contrast medium IV injection
Start delay
exam dependent
Flow rate
dependent upon needle size/Access site
Total amount
1 – 2 ml per kg of body weight
* The mAs should be adjusted to the age of the child.
* The conversion factor for an 8-week-old,
and a scan range of 90 mm was used.
** The conversion factor for a 7-year-old, 260 mAs
and a scan range of 110 mm was used.
Body Weight kV
mAs
CTDI
w
Effective dose
(mGy)
(mSv)
< 6 months
120 90
20.8
Male: 2.1*
Female: 2.1*
6 months –
120 150
34.8
3 years
3 years –
120 220
51.0
Male: 5.4**
6 years
Female: 5.2**