AbdThinSlice
2
nd
reconstr.
kV
120
Effective mAs
140
Slice collimation
0.75 mm
Slice width
5.0 mm
1.0 mm
Feed/Rotation
12.0 mm
Rotation time
0.5 sec.
Kernel
B30f
B20f
Increment
5.0 mm
0.7 mm
Image order
cr-ca
CTDI
w
10.9 mGy
Effective dose
Male: 3.4 mSv
Female: 4.1 mSv
If you want to use this protocol for a two phase study,
repeat the same protocol as mentioned below, and
chose start delay time for arterial phase as 20-25 s. In
this case, the thin slice reconstruction can also be used
for postprocessing. But do not administer oral contrast
medium, as this impairs the editing of MIP/SSD/VRT
images. Water could be used instead if necessary.
You can repeat the same protocol simply by clicking
the chronicle with the right mouse button for “repeat“.
E.g. when both non-contrast and contrast studies are
required.
MPRthick: AbdThinSlice
coronal
Image thickness
3
Distance between images
3
Number of images
50
The 2
nd
reconstruction can be loaded in the 3D Card,
MPR. Select the Range function and a predefined MPR
range, linked to the series, will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
Contrast medium IV injection
Start delay
50 – 65 sec.
Flow rate
4 ml/sec.
Total amount
100 ml
119
Abdomen
Содержание SOMATOM Sensation 16
Страница 1: ...SOMATOM Sensation 16 Application Guide Routine Protocols Software Version A50 ...
Страница 7: ...7 Content ...
Страница 23: ...23 General ...
Страница 29: ...29 Children ...
Страница 49: ...49 Children ...
Страница 55: ...55 Children ...
Страница 94: ...94 Shoulder Overview ShoulderSpi Spiral mode for HiRes bone studies and soft tissues ...