
Head
AngioThinSlice
2
nd
reconstr.
kV
120
Effective mAs
130
Slice collimation
0.75 mm
Slice width
4.0 mm
0.75 mm
Feed/Rotation
6.8 mm
Rotation time
0.5 sec.
Kernel
H20f
H10f
Increment
4.0 mm
0.5 mm
Image order
ca-cr
CTDI
w
27.4 mGy
Effective dose
Male: 0.6 mSv
Female: 0.7 mSv
Use of CARE Bolus with monitoring scans positioned at
the level of the basilar artery or carotid artery. Set the
trigger threshold at 120 HU, or use manual triggering.
MPRthick: AngioThinSlice
coronal
Image thickness
3
Distance between images
3
Number of images
35
For the 2nd reconstruction the Autoload into MPRthick
Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MPRthick, and a coronal
MPRthick Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
Contrast medium IV injection
Start delay
18 sec.
Flow rate
3.5 ml/sec.
Total amount
75 ml
75
Содержание SOMATOM Sensation 16
Страница 1: ...SOMATOM Sensation 16 Application Guide Routine Protocols Software Version A50 ...
Страница 7: ...7 Content ...
Страница 23: ...23 General ...
Страница 29: ...29 Children ...
Страница 49: ...49 Children ...
Страница 55: ...55 Children ...
Страница 94: ...94 Shoulder Overview ShoulderSpi Spiral mode for HiRes bone studies and soft tissues ...
Страница 115: ...115 Abdomen ...