Head
ThinSliceSeq
RoutineSeq
kV
120
120
Effective mAs
260
260
Slice collimation 0.75 mm
1.5 mm
Slice width
4.5 mm
9.0 mm
Feed/Scan
9.0 mm
18.0 mm
Rotation time
1.0 sec.
1.0 sec.
Kernel
H40s
H40s
Image order
ca-cr
ca-cr
CTDI
w
58.2 mGy
60.3 mGy
Effective dose
Male: 0.7 mSv
Male: 1.6 mSv
Female: 0.8 mSv Female: 1.8 mSv
PosteriorFossaOptimization Filter
PFO: To reduce beam-hardening artifacts in head
images, particularly in the base of the skull, use the
Posterior Fossa Optimization (PFO) filter.
If you want to set up the scan with only one range,
i. e. either 12 x 0.75 mm or 12 x 1.5 mm, you can
easily delete one mode by clicking the chronicle with
the right mouse button, and select cut.
Contrast medium IV injection
Start delay
60 sec.
Flow rate
2 ml/sec.
Total amount
50 – 60 ml
77