147
ExtrUHRSpi
kV
120
Effective mAs
120
Slice collimation
0.75 mm
Slice width
0.75 mm
Feed/Rotation
6.8 mm
Rotation time
0.75 sec.
Kernel
U90u
Increment
0.5 mm
Image order
cr-ca
CTDI
w
10.1 mGy
Effective dose
Male: 0.01 mSv
Female: 0.01 mSv
MPRthick: ExtrUHRSpi
coronal
Image thickness
2
Distance between images
2
Number of images
30
This protocol is used for ultra high-resolution
studies. It is not recommended to use this protocol for
3D studies.
For image reconstruction of soft tissue, use kernel
U30.
The 2
nd
reconstruction can be loaded in the 3D Card,
MPR. Select the Range function and a predefined range,
linked to the series, will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.