137
Spine
ThinSliceSpi
Indications:
Thin slice spiral mode for spine studies when high
resolution for fine structure is required.
ThinSliceSpi
2
nd
reconstr.
kV
120
Effective mAs
330
Slice collimation
0.75 mm
Slice width
3.0 mm
1.0 mm
Feed/Rotation
6.0 mm
Rotation time
0.75 sec.
Kernel
B60s
B60s
Increment
3.0 mm
0.6 mm
Image order
cr-ca
CTDI
w
27.7 mGy
Effective dose
Male: 6.6 mSv
Female: 13.2 mSv
MPRthick: ThinSliceSpi
sagittal
Image thickness
3
Distance between images
3
Number of images
20
For the 2
nd
reconstruction the Autoload into the MPR
Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MPR, and a coronal
MPRthick Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.