LungLowDose
2
nd
reconstr.
kV
120
Effective mAs
20
Slice collimation
0.75 mm
Slice width
5.0 mm
1.0 mm
Feed/Rotation
18.0 mm
Rotation time
0.5 sec.
Kernel
B50f
B20f
Increment
5.0 mm
0.7 mm
Image order
cr-ca
CTDI
w
1.6 mGy
Effective dose
Male: 0.6 mSv
Female: 0.7 mSv
101
You could repeat the same protocol simply by clicking
the chronicle with the right mouse button for “repeat“.
E.g. when both non-contrast and contrast studies are
required.
For lung cancer staging, this protocol can be combined
with protocol “Neck Routine“.
Contrast medium IV injection
Start delay
30 sec.
Flow rate
2.5 ml/sec.
Total amount
50 – 70 ml
Thorax