SITRANS LG240 - Operating Instructions
44286-EN-171021
Specifications apply also to the current output
8)
Temperature drift - Current output
±0.03 %/10 K relating to the 16 mA span max. ±0.3 %
Deviation in the current output due to digital/analogue conversion
Ʋ Non-Ex and Ex-ia version
< ±15 µA
Ʋ Ex-d-ia version
< ±40 µA
Additional deviation through electromag-
netic interference acc. to EN 61326
< ±150 µA
Influence of the superimposed gas and pressure on measurement accuracy
The propagation speed of the radar impulses in gas or vapour above the medium is reduced by
high pressure. This effect depends on the superimposed gas or vapours.
The following table shows the resulting deviation for some typical gases and vapours. The specified
values refer to the distance. Positive values mean that the measured distance is too large, negative
values that the measured distance is too small.
Gas phase
Temperature
Pressure
1 bar (14.5 psig)
10 bar (145 psig)
50 bar (725 psig)
Air
20 °C (68 °F)
0 %
0.22 %
1.2 %
200 °C (392 °F)
-0.01 %
0.13 %
0.74 %
400 °C (752 °F)
-0.02 %
0.08 %
0.52 %
Hydrogen
20 °C (68 °F)
-0.01 %
0.1 %
0.61 %
200 °C (392 °F)
-0.02 %
0.05 %
0.37 %
400 °C (752 °F)
-0.02 %
0.03 %
0.25 %
Steam (saturated
steam)
100 °C (212 °F)
0.26 %
-
-
180 °C (356 °F)
0.17 %
2.1 %
-
264 °C (507 °F)
0.12 %
1.44 %
9.2 %
366 °C (691 °F)
0.07 %
1.01 %
5.7 %
Characteristics and performance data
Measuring cycle time
< 500 ms
Step response time
9)
≤ 3 s
Max. filling/emptying speed
1 m/min
Products with high dielectric constant (>10) up to 5 m/
min.
Ambient conditions
Ambient, storage and transport tempera-
ture
-40 … +80 °C (-40 … +176 °F)
8)
Also for the additional current output (optional).
9)
Time span after a sudden measuring distance change by max. 0.5 m in liquid applications, max 2 m with bulk
solids applications, until the output signal has taken for the first time 90 % of the final value (IEC 61298-2).
Содержание SITRANS LG240
Страница 2: ...2 PBD 51041038 SITRANS LG240 Operating Instructions 44286 EN 171021 ...
Страница 85: ...Notes 85 PBD 51041038 SITRANS LG240 Operating Instructions 44286 EN 171021 ...
Страница 86: ...Notes 86 PBD 51041038 SITRANS LG240 Operating Instructions 44286 EN 171021 ...
Страница 87: ...Notes 87 PBD 51041038 SITRANS LG240 Operating Instructions 44286 EN 171021 ...