71
PBD-51041038
SITRANS LG240 - Operating Instructions
44286-EN-171021
2
4
5
3
1
Fig. 39: Measuring ranges - SITRANS LG240
1 Reference plane
2 Probe length L
3 Measuring range (default setting refers to the measuring range in water)
4 Upper dead band (see following diagrams - grey section)
5 Lower dead band (see following diagrams - grey section)
Typical deviation - Interface measure-
ment
± 5 mm (0.197 in)
Typical deviation - Total level interface
measurement
See following diagrams
Typical deviation - Level measurement
6)7)
See following diagrams
6)
Depending on the mounting conditions, deviations can occur which can be rectified by adapting the adjustment
or changing the measured value offset in the DTM service mode.
7)
The dead bands can be optimized via a false signal suppression.