SITRANS LG240 - Operating Instructions
44286-EN-171021
Measurement error with constant level
Fault description
Error pattern
Cause
Rectification
1. Measured value
shows a too low or too
high level
Leve
l
time
0
•
Min./max. adjustment not
correct
•
Adapt min./max. adjustment
•
Incorrect linearisation curve
•
Adapt linearisation curve
•
Running time error (small meas-
urement error close to 100 %/
serious error close to 0 %)
•
Repeat setup
2. Measured value
jumps towards 100 %
Level
time
0
•
Due to the process, the
amplitude of the product echo
decreases
•
A false signal suppression was
not carried out
•
Carry out a false signal suppres-
sion
•
Amplitude or position of a
false signal has changed (e.g.
buildup); false signal suppres-
sion no longer matches
•
Determine the reason for the
changed false signals, carry out
false signal suppression, e.g.
with buildup
Measurement error during filling
Fault description
Error pattern
Cause
Rectification
3. Measured value re-
mains in the area of the
bottom during filling
Leve
l
time
0
•
Echo from the probe end larger
than the product echo, for exam-
ple, with products with ε
r
< 2.5
oil-based, solvents, etc.
•
Check parameter "Medium"
and "Vessel height", adapt if
necessary
4. Measured value re-
mains momentarily
unchanged during fill-
ing and then jumps to
the correct level
Level
time
0
•
Turbulence on the product
surface, quick filling
•
Check parameters, change if
necessary, e.g. in dosing vessel,
reactor
5. Measured value
jumps sporadically to
100 % during filling
Leve
l
time
0
•
Changing condensation or
contamination on the probe
•
Carry out a false signal suppres-
sion
6. Measured value
jumps to ≥ 100 % or
0 m distance
Leve
l
time
0
•
Level echo is no longer detected
in the close range due to false
signals in the close range. The
sensor goes into overfill protec-
tion mode. The max. level (0 m
distance) as well as the status
message "Overfill protection"
are outputted.
•
Eliminate false signals in the
close range
•
Check installation conditions
•
If possible, switch off the func-
tion "Overfill protection"
Содержание SITRANS LG240
Страница 2: ...2 PBD 51041038 SITRANS LG240 Operating Instructions 44286 EN 171021 ...
Страница 85: ...Notes 85 PBD 51041038 SITRANS LG240 Operating Instructions 44286 EN 171021 ...
Страница 86: ...Notes 86 PBD 51041038 SITRANS LG240 Operating Instructions 44286 EN 171021 ...
Страница 87: ...Notes 87 PBD 51041038 SITRANS LG240 Operating Instructions 44286 EN 171021 ...