
1010BXFM-3
Section 4
4.5.1 THE ANALOG OUTPUT TRIM MENU
Analog Output Trim allows you to fine-tune the meter’s analog voltage and current outputs
using a multimeter connected to the output under test. In addition, you can use a frequency
counter to fine-tune the meter’s pulse rate output.
NOTE: The current and Pgen trimming will be limited by the 12-bit resolution of the
meter’s D/A Convertor (DAC).
To access the [Analog Output Trim] menu press
4.5.2 CURRENT OUTPUT TRIM (Io1 & Io2)
(Note: Can be trimmed to within .005 mA of nominal.)
To calibrate a current output (Io1 or Io2):
Set up the multimeter to read Amps, then connect it to the supply and return term-
inals of the current output under test.
Move the highlight to the port to be tested, press , then press to move the cursor
to Trim @ 4mA.
Press . This triggers a 4.00 mA pop-up window. The multimeter should now be
reading 4.00 mA.
If the multimeter reading does not match, use PC numeric keys or Magnetic Wand to
type in the multi-meter reading.
Press to register setting. This adjusts the meter’s DAC (digital-to-analog con-
verter) so that a 4mA output corresponds with 4mA on the multimeter.
Re-check the multimeter to make sure that it is now reading 4mA.
4.5.3 PGEN OUTPUT TRIM (Single Channel models only)
(Note: Can be trimmed to within 1.25 Hz of nominal.)
To calibrate a pulse rate output (Pgen1):
Connect a frequency counter to the supply and return terminals of the pulse rate
output under test.
Move the highlight to the port to be tested , press the , then press
to move the cursor to Trim @ 1kHz.
Press . This triggers a 1 kHz pop-up window. The frequency counter should now
be reading 1 kHz.
If the frequency counter reading does not match, use PC numeric keys or Magnetic
Wand to enter the frequency counter reading.
ENT
ENT
ANALOG OUTPUT TRIM MENU STRUCTURE
Analog Output Trim
Io1
Operate
Trim @ 4mA
Indicated mA = x.xx
Io2
Operate
Trim @ 4mA
Indicated mA = x.xx
Pgen
Operate (Single Channel only)
Trim @ 1kHz
Indicated Hz = xxxx
Ö
Ú
Ö
Ú
Ö
Ú
Ö
Ö
Ö
Ö
Ö
Ú
Ö
ENT
Содержание SITRANS FUH1010PVX
Страница 2: ......
Страница 6: ......
Страница 14: ......
Страница 28: ......
Страница 130: ......
Страница 186: ......
Страница 187: ...Siemens Hardware ...
Страница 188: ......
Страница 189: ...Siemens Hardware ...
Страница 190: ......
Страница 191: ...Siemens Hardware ...
Страница 192: ......
Страница 193: ...Siemens Hardware ...
Страница 194: ......
Страница 195: ...Siemens Hardware ...
Страница 196: ...1010N 5 7 INSTALLATION DRAWING ANALOG INPUT MODULE 21614 C ...
Страница 200: ...1011HNS 7 USE OF FLEXIBLE CONDUIT WITH TRANSDUCERS SIZE A B 21614 C USE OF FLEXIBLE CONDUIT WITH TRANSDUCERS SIZE C D ...
Страница 201: ...1011HNS 8 TRANSDUCER PART NUMBER L H W SIZE C D SIZE A B 21614 C Siemens Hardware ...
Страница 202: ......
Страница 203: ...Siemens Hardware ...
Страница 204: ......
Страница 206: ......
Страница 214: ......
Страница 225: ...Siemens Hardware ...