1010BXFM-3
Section 2
996PSP-0
0
2.420 PVC 0.076
Direct A 6
996PSP-1
1
2.481 PVC 0.077
Direct A 9
996PSP-2
2
2.743 PVC 0.136
Direct A 6
996PSP-3
3
3.758 PVC 0.344
Direct A 1
996PSP-4
4
6.500 PVC 1.000
Direct** A 0
NOTE: When using a 996PS-4 Simulator, you must access the [Application Data] menu,
select [Liquid Type] - [Other], then program the [Estimated Vs] for a value of
2100 m/s.
Access the [Pick/Install Xdcr] menu. Access the [Transducer Model] menu cell. Enter
the Transducer Model option list. Select [991 Universal].
Check the [Transducer Size] menu cell. If necessary, enter the Transducer Size option
list and pick the actual transducer under test.
Check the [Xdcr Mount Mode] menu cell. Adjust as necessary to match the simulator
chart above.
Move the highlight down to [Install Completed?]. Access the option list. Move the cursor
to [Install]. Press <ENT> to start the transducer install routine.
Check to make sure that the computer returns the Number Index listed in the chart above.
Next, review and verify the Diagnostics data items as described previously. If the com-
puter, cables and transducers are operating properly, then review all site conditions to
locate the operating problem.
If a Pipe Simulator/Test-Block Test Fails:
Replace the transducer cables. If this allows you to complete the test as described above,
then the cables were defective.
If the cables are proven to be good, replace the transducers on the pipe simulator with a
“known good” set. If the system functions properly, then the original transducers under
test are defective. Please return any defective transducers to Controlotron for repair.
If you replace the transducers and cables, and the system still fails to function correct-
ly, then it is likely that the computer has a malfunctioning circuit module. If you have
spare modules you may try module substitution to identify the defective module. Other
wise, please call Controlotron’s Technical Service Department for further instructions
Part Xdcr Pipe Pipe Wall Mount Letter Number
Number Size OD (in) Mat’l Thk (in) Mode Index Index*
996PSP PIPE SIMULATOR CHART (English)
* The computer generates the Number Index.
** See note below.
* The computer generates the Number Index.
Part Xdcr Pipe Pipe Wall Mount Letter Number
Number Size OD (mm) Mat’l Thk (mm) Mode Index Index*
996PSP-0
0
61.47 PVC 1.93
Direct A 6
996PSP-1
1
63.02 PVC 1.96
Direct A 9
996PSP-2
2
69.67 PVC 3.45
Direct A 6
996PSP-3
3
95.45 PVC 8.74
Direct A 1
996PSP-4
4
165.1 PVC 25.4
Direct** A 0
996PSP PIPE SIMULATOR CHART (Metric)
** See note below.
Содержание SITRANS FUH1010PVX
Страница 2: ......
Страница 6: ......
Страница 14: ......
Страница 28: ......
Страница 130: ......
Страница 186: ......
Страница 187: ...Siemens Hardware ...
Страница 188: ......
Страница 189: ...Siemens Hardware ...
Страница 190: ......
Страница 191: ...Siemens Hardware ...
Страница 192: ......
Страница 193: ...Siemens Hardware ...
Страница 194: ......
Страница 195: ...Siemens Hardware ...
Страница 196: ...1010N 5 7 INSTALLATION DRAWING ANALOG INPUT MODULE 21614 C ...
Страница 200: ...1011HNS 7 USE OF FLEXIBLE CONDUIT WITH TRANSDUCERS SIZE A B 21614 C USE OF FLEXIBLE CONDUIT WITH TRANSDUCERS SIZE C D ...
Страница 201: ...1011HNS 8 TRANSDUCER PART NUMBER L H W SIZE C D SIZE A B 21614 C Siemens Hardware ...
Страница 202: ......
Страница 203: ...Siemens Hardware ...
Страница 204: ......
Страница 206: ......
Страница 214: ......
Страница 225: ...Siemens Hardware ...