3-15
1010BXFM-3
Section 3
INSTALLATION - REFLECT MOUNT WITH TRANSDUCERS AND SPACING BAR
1. Perform all required menu steps until the flow computer issues the number index and
prompts you to press <ENT> to finish the transducer install routine. Stop at this point.
Note the number index value displayed in the Pick/Install menu. You will use this index to
properly space the transducers. Check to ensure that you have a matched set of transducers.
They both should have the same S/N number but marked with either an “A” or “B” (e.g., 100A
and 100B).
2. Assemble the transducers to the spacer bar, with the cable connectors facing away from
each other as shown above. The spacer bar is attached to a transducer using a transducer
index screw. One transducer is attached using the “REF” hole on the spacer bar. The sec-
ond transducer is attached to the spacer at the index hole specified in Step 1.
3. Temporarily position the assembly (in the 9 o’clock position) at the location where you
have determined it would be mounted. Ensure that it is a smooth area without any raised
spots or seams. Mark a generous area around the transducers (1/2-inch on either side and
half again the length front and back) with a pencil or chalk. Remove the assembly.
4. Prepare the two areas you marked by de-greasing the surface, if needed, and removing any
grit, corrosion, rust, loose paint or surface irregularities with the abrasive pipe condition-
ing material provided. Clean the pipe of all debris and abrasive particles.
5. Remove transducer from the spacer bar that was attached through the REF hole. Put a
mounting strap around the pipe and engage an end into the adjusting screw (adjusting
screw should be pointing up). Apply a 1/8-inch continuous bead of couplant compound down
the center (the long way) of the contact surface of one of the transducers (see paragraph
3.3.5 Recommended Sonic Coupling Compounds). Place the transducer on the pipe center
in the middle of one of the areas you have cleaned with its cable connector pointing away
from the other cleaned area. Holding the transducer in place, slide the mounting strap over
it (and under the clip if there is one) and tighten with a screwdriver. While tightening,
check to ensure that the white dot under the front label is centered on the pipe and that
there is equal space on both edges. Also, Make sure to position strap for easy access to the
mounting strap adjusting screw.
6. Repeat Step 5 with the second transducer leaving it still attached to the spacer bar. Apply
a 1/8-inch bead of couplant to the transducer. At the same time you place it in the middle of
Mounting
Strap
Spacer Bar
Number Index
Securing
Screw
1011 Series
Transducer
Spacer Bar
Spring Clip
Spacer Bar Reference
Index Securing Screw
Содержание SITRANS FUH1010PVX
Страница 2: ......
Страница 6: ......
Страница 14: ......
Страница 28: ......
Страница 130: ......
Страница 186: ......
Страница 187: ...Siemens Hardware ...
Страница 188: ......
Страница 189: ...Siemens Hardware ...
Страница 190: ......
Страница 191: ...Siemens Hardware ...
Страница 192: ......
Страница 193: ...Siemens Hardware ...
Страница 194: ......
Страница 195: ...Siemens Hardware ...
Страница 196: ...1010N 5 7 INSTALLATION DRAWING ANALOG INPUT MODULE 21614 C ...
Страница 200: ...1011HNS 7 USE OF FLEXIBLE CONDUIT WITH TRANSDUCERS SIZE A B 21614 C USE OF FLEXIBLE CONDUIT WITH TRANSDUCERS SIZE C D ...
Страница 201: ...1011HNS 8 TRANSDUCER PART NUMBER L H W SIZE C D SIZE A B 21614 C Siemens Hardware ...
Страница 202: ......
Страница 203: ...Siemens Hardware ...
Страница 204: ......
Страница 206: ......
Страница 214: ......
Страница 225: ...Siemens Hardware ...