Influence variables
Sample gas pressure
Deviation at zero point
≤ 0.2 % of smallest measuring range / 1% pressure
change or ≤ 50 vpm O
2
/ 1% pressure change,
whichever is larger
Deviation from the span
gas
Deviation ≤ 0.2% of current span / 1% pressure
change or 50 vpm O
2
/ 1% pressure change, which‐
ever is larger
Sample gas flow
Deviation at zero point
≤ 1 % of smallest measuring range per 0.1 l / min
flow rate change or 50 vpm O
2
per 0.1 l / min flow
rate change within the permissible flow rate range
(The larger value is always in effect)
Deviation from the span
gas
≤ 1 % of smallest measured value per 0.1 l / min
flow rate change or 50 vpm O
2
per 0.1 l / min flow
rate change within the permissible flow rate range
(The larger value is always in effect)
Accompanying gases
Zero point deviation (cross-sensitivity) in accord‐
ance with Table A.1 of EN 61207-3
Power supply (mains voltage variations in range
85 to 264 V AC / 47 to 63 Hz)
≤ ± 0.1% of full-scale value
Electrical inputs and outputs
Analog and digital interfaces
See Operating Instructions of field control unit
Table 10-15 Climatic conditions
Climatic conditions
Permissible ambient temperature
● -30 to +70 °C during storage and transportation
● +5 °C ... +55 °C for regular operation
● +5 °C ... +60 °C with limited measuring
accuracy
Ambient humidity (rel. humidity) during storage,
transportation or operation
< 90 % (condensation on the installed components
must be avoided)
Table 10-16 Gas connections
Gas connections
With pipes
Connection fitting for pipe 6 mm diameter
Table 10-17 Parts in contact with the sample gas
Material of parts in contact with the sample gas
Measuring chamber
Stainless steel: Mat. no. 1.4571
Gas path
Pipes (sample gas), including flange: Stainless
steel Mat. no. 1.4404
Technical specifications
10.3 OXYMAT 7 technical specifications
Field device
92
Operating Instructions, 05/2015, A5E35640463-02
Содержание SIPROCESS GA700
Страница 6: ...Inhaltsverzeichnis Feldgerät 6 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 22: ...Allgemeine Sicherheitshinweise 2 9 Automatisches Reset Feldgerät 22 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 64: ...Anschließen 6 5 Feldbediengerät Feldgerät 64 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 66: ...Bedienen 7 1 Bedienen Feldgerät 66 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 94: ...Technische Daten 10 3 Technische Daten OXYMAT 7 Feldgerät 94 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 112: ...Index Feldgerät 112 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 118: ...Table of contents Field device 6 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 134: ...General safety instructions 2 9 Automatic reset Field device 22 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 158: ...Installing mounting 5 3 Field module Field device 46 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 176: ...Wiring 6 5 Field control unit Field device 64 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 178: ...Operation 7 1 Operation Field device 66 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 190: ...Commissioning 8 4 Commissioning the SIPROCESS GA700 Field device 78 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 222: ...Appendix A 6 Returned delivery Field device 110 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 224: ...Index Field device 112 Operating Instructions 05 2015 A5E35640463 02 ...