8.3
OXYMAT 7
WARNING
Incorrect reference gas
Explosion hazard when used in the hazardous area.
Use nitrogen exclusively as the reference gas for combustible media in the hazardous area.
Also ensure that the reference gas pressure is at least 2.2 bar (3.2 bar abs.).
8.3.1
Reference gas
Only used nitrogen as the reference gas.
Make sure that the reference gas pressure is 2000 hPa +/- 150 hPa above the sample gas
pressure.
8.3.2
Zero point error
Accompanying gases (interfering gases) cause a zero point deviation that is positive with
paramagnetic gases and negative with diamagnetic gases. The extent of this offset per 100
vol. % accompanying gas, expressed in vol. % O
2
absolute, is listed in the following table. The
values in the tables refer to nitrogen at 60 °C and 1 000 hPa absolute to IEC 1207/3.
Table 8-1
Zero point error of organic gases
Accompanying gas (concentration 100 vol. %)
Zero point deviation in
vol. % O
2
absolute
Ethane
C
2
H
6
-0.49
Ethene (ethylene)
C
2
H
4
-0.22
Ethine (acetylene)
C
2
H
2
-0.29
1.2 butadiene
C
4
H
6
-0.65
1.3 butadiene
C
4
H
6
-0.49
n-butane
C
4
H
10
-1.26
Isobutane
C
4
H
10
-1.30
1-butene
C
4
H
8
-0.96
Isobutene
C
4
H
8
-1.06
Dichlorodifluoromethane (R12)
CCl
2
F
2
-1.32
Ethanoic acid
CH
3
COOH
-0.64
n-heptane
C
7
H
16
-2.40
n-hexane
C
6
H
14
-2.02
Cyclo-hexane
C
6
H
12
-1.84
Methane
CH
4
-0.18
Methanol
CH
3
OH
-0.31
Commissioning
8.3 OXYMAT 7
Field device
70
Operating Instructions, 05/2015, A5E35640463-02
Содержание SIPROCESS GA700
Страница 6: ...Inhaltsverzeichnis Feldgerät 6 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 22: ...Allgemeine Sicherheitshinweise 2 9 Automatisches Reset Feldgerät 22 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 64: ...Anschließen 6 5 Feldbediengerät Feldgerät 64 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 66: ...Bedienen 7 1 Bedienen Feldgerät 66 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 94: ...Technische Daten 10 3 Technische Daten OXYMAT 7 Feldgerät 94 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 112: ...Index Feldgerät 112 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 118: ...Table of contents Field device 6 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 134: ...General safety instructions 2 9 Automatic reset Field device 22 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 158: ...Installing mounting 5 3 Field module Field device 46 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 176: ...Wiring 6 5 Field control unit Field device 64 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 178: ...Operation 7 1 Operation Field device 66 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 190: ...Commissioning 8 4 Commissioning the SIPROCESS GA700 Field device 78 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 222: ...Appendix A 6 Returned delivery Field device 110 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 224: ...Index Field device 112 Operating Instructions 05 2015 A5E35640463 02 ...