5. Define measuring ranges:
[2.03] Settings > Measuring ranges
Note
Putting back into operation with disabled measuring ranges
The measuring ranges are activated in the factory. If you disable the measuring ranges
(setting: "Off"), all the setting options in other menus related to the measuring range are
also hidden. The measuring ranges can no longer be used.
If you put the device back into operation with fully deactivated measuring ranges, the status
bar of the display shows message 150 "Parameter value invalid". Before you continue to
put the device back into operation:
● Activate at least one measuring range.
● Disable the autoranging function if it is enabled.
6. Set inputs/outputs: [2.09] Inputs/outputs
– Set digital inputs: [2.09.3] Digital inputs
– Setting digital outputs [2.09.5] Digital outputs
If you have the appropriate optional modules:
– Set analog inputs: [2.09.1] Setting analog inputs
– Set analog outputs: [2.09.2] Setting analog outputs
7. Check setting of gas path and change, if necessary:
[2.12] Settings > Gas path/process tag label
8. Calibrate the device: [2.08] Settings > Calibrations
– Familiarize yourself with the basics of calibration. Calibration changes internal device
characteristics.
– Prepare for the calibration. Set the setpoints, calibration tolerances, and the sample gas
purging time: [2.08.1] Setpoints, tolerances, purging time
– Assign parameters for a calibration process: [2.08.3] Execute calibrations
9. Save parameter set: [3.10] Maintenance & Diagnostics > Save/load parameter set
Approach in case of an incomplete/faulty boot process
If the device detects a serious error during booting, the boot process is aborted. The start
screen is no longer updated. In this case, contact the service technician responsible.
See also
Commissioning
8.4 Commissioning the SIPROCESS GA700
Field device
Operating Instructions, 05/2015, A5E35640463-02
77
Содержание SIPROCESS GA700
Страница 6: ...Inhaltsverzeichnis Feldgerät 6 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 22: ...Allgemeine Sicherheitshinweise 2 9 Automatisches Reset Feldgerät 22 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 64: ...Anschließen 6 5 Feldbediengerät Feldgerät 64 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 66: ...Bedienen 7 1 Bedienen Feldgerät 66 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 94: ...Technische Daten 10 3 Technische Daten OXYMAT 7 Feldgerät 94 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 112: ...Index Feldgerät 112 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 118: ...Table of contents Field device 6 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 134: ...General safety instructions 2 9 Automatic reset Field device 22 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 158: ...Installing mounting 5 3 Field module Field device 46 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 176: ...Wiring 6 5 Field control unit Field device 64 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 178: ...Operation 7 1 Operation Field device 66 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 190: ...Commissioning 8 4 Commissioning the SIPROCESS GA700 Field device 78 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 222: ...Appendix A 6 Returned delivery Field device 110 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 224: ...Index Field device 112 Operating Instructions 05 2015 A5E35640463 02 ...