
oxygen partial pressure). When you combine two gases with differing oxygen content in a
magnetic field, a partial pressure difference arises between them. The measuring effect is
therefore always based on the difference in the oxygen content of the two gases. The two
gases are referred to as the sample gas and the reference gas.
To measure oxygen in the OXYMAT 7, the reference gas (nitrogen, air or oxygen) flows through
two reference gas channels
③
into the measuring chamber
⑥
. One of these partial flows
enters the measuring chamber in the area of the magnetic field
⑦
. If the sample gas contains
no oxygen, the reference gas flows freely into the sample chamber. If the sample gas does
contain oxygen, the oxygen molecules concentrate in the area of the magnetic field. The
reference gas can then no longer flow freely into the sample chamber. There is a slight build-
up of dynamic pressure, which depends on the concentration of oxygen in the sample gas. In
an oscillating magnetic field, oscillating pressure is also generated and this results in
alternating flow. This alternating flow is converted into an alternating current signal by two
microflow sensors
④
located between the reference gas channels
③
.
The microflow sensors consist of two nickel-plated grids each, which are arranged together
with two further nickel-plated grids to form a measuring bridge. The alternating flow results in
a change in the resistance of the nickel-plated grids. The resulting offset in the bridge is a
measure of the concentration of oxygen in the sample gas.
The microflow sensors
④
are surrounded by reference gas. This arrangement protects them
from corrosion by the sample gas.
Vibrations at the installation site can interfere with the measured signal, for example, large
fluctuations in the output signal. Two additional optional microflow sensors
⑩
serve as
vibration sensors and thus compensate for the undesirable behavior.
Description
3.6 Principle of operation of OXYMAT 7
Field device
Operating Instructions, 05/2015, A5E35640463-02
33
Содержание SIPROCESS GA700
Страница 6: ...Inhaltsverzeichnis Feldgerät 6 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 22: ...Allgemeine Sicherheitshinweise 2 9 Automatisches Reset Feldgerät 22 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 64: ...Anschließen 6 5 Feldbediengerät Feldgerät 64 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 66: ...Bedienen 7 1 Bedienen Feldgerät 66 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 94: ...Technische Daten 10 3 Technische Daten OXYMAT 7 Feldgerät 94 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 112: ...Index Feldgerät 112 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 118: ...Table of contents Field device 6 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 134: ...General safety instructions 2 9 Automatic reset Field device 22 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 158: ...Installing mounting 5 3 Field module Field device 46 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 176: ...Wiring 6 5 Field control unit Field device 64 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 178: ...Operation 7 1 Operation Field device 66 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 190: ...Commissioning 8 4 Commissioning the SIPROCESS GA700 Field device 78 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 222: ...Appendix A 6 Returned delivery Field device 110 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 224: ...Index Field device 112 Operating Instructions 05 2015 A5E35640463 02 ...