9.4
Errors
Table 9-1
Error messages
Fault message
Description
Measures
Invalid device
configuration
No analyzer module was detected. At least one analyzer module is re‐
quired to use the device.
Install at least one analyzer
module.
Firmware ver‐
sions incompati‐
ble
The firmware of the basic device and the analyzer modules must be com‐
patible with one another.
Please contact the service de‐
partment.
Unknown mod‐
ule type detected
An analyzer module has been detected which is not supported in the
previous firmware version.
Please contact the service de‐
partment.
Memory error
detected
A hardware fault, faulty firmware or no firmware at all has been detected
in the following analyzer modules, e.g., AM1. Possible causes:
● No firmware is present on the analyzer module.
● The existing firmware is invalid.
● The program memory is faulty.
In this case, the analyzer module starts in bootloader mode and is ready
to perform a firmware update.
If the error still exists despite the firmware update, the analyzer module
PCB must be replaced.
Please contact the service de‐
partment.
Cable fault de‐
tected
The installed analyzer modules could not be recognized correctly.
Possible causes: Faulty or incorrectly connected CAN bus cable.
Check the CAN bus connec‐
tions.
Service and maintenance
9.4 Errors
Field device
Operating Instructions, 05/2015, A5E35640463-02
85
Содержание SIPROCESS GA700
Страница 6: ...Inhaltsverzeichnis Feldgerät 6 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 22: ...Allgemeine Sicherheitshinweise 2 9 Automatisches Reset Feldgerät 22 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 64: ...Anschließen 6 5 Feldbediengerät Feldgerät 64 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 66: ...Bedienen 7 1 Bedienen Feldgerät 66 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 94: ...Technische Daten 10 3 Technische Daten OXYMAT 7 Feldgerät 94 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 112: ...Index Feldgerät 112 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 118: ...Table of contents Field device 6 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 134: ...General safety instructions 2 9 Automatic reset Field device 22 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 158: ...Installing mounting 5 3 Field module Field device 46 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 176: ...Wiring 6 5 Field control unit Field device 64 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 178: ...Operation 7 1 Operation Field device 66 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 190: ...Commissioning 8 4 Commissioning the SIPROCESS GA700 Field device 78 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 222: ...Appendix A 6 Returned delivery Field device 110 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 224: ...Index Field device 112 Operating Instructions 05 2015 A5E35640463 02 ...