6.3
Field module
6.3.1
Gas connections
Safety instructions
WARNING
Possible leakages of gas path
Fit the gas connections in accordance with the mounting specifications of the manufacturer
using a suitable open-ended wrench. In the process, make sure that you counter properly.
Otherwise, there is a danger that the gas path will leak.
WARNING
Danger of poisoning by escaping gas
Components which come into contact with the sample gas must be resistant to it. Otherwise,
there is a danger that the gas path will leak and that sample gas will accumulate in the
analyzer.
Take appropriate precautions to prevent gas from escaping, e.g. by monitoring the flow
outside the analyzer.
Since it is not possible to absolutely guarantee that toxic or corrosive gases cannot escape
from or accumulate in the enclosure - even if to a limited extent - the field module must be
purged with oil-free and dry purging air (e.g. from the non-hazardous area) or with inert gas
when measuring corrosive or toxic gases.
Calculation of the required purging gas flow must be carried out by a technical expert. The
discharged purging air must be disposed of environmentally-friendly via a suitable exhaust
line.
An additional flame arrester to which the exhaust line is connected must be provided in this
case in addition to the pressure compensation connection.
Note
Reference gas connection
The reference gas connection is designed as a flame arrester. The gas supply is via a 6 mm
pipe, and the connection is made using a clamping ring.
Only use nitrogen as the reference gas.
Wiring
6.3 Field module
Field device
50
Operating Instructions, 05/2015, A5E35640463-02
Содержание SIPROCESS GA700
Страница 6: ...Inhaltsverzeichnis Feldgerät 6 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 22: ...Allgemeine Sicherheitshinweise 2 9 Automatisches Reset Feldgerät 22 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 64: ...Anschließen 6 5 Feldbediengerät Feldgerät 64 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 66: ...Bedienen 7 1 Bedienen Feldgerät 66 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 94: ...Technische Daten 10 3 Technische Daten OXYMAT 7 Feldgerät 94 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 112: ...Index Feldgerät 112 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 118: ...Table of contents Field device 6 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 134: ...General safety instructions 2 9 Automatic reset Field device 22 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 158: ...Installing mounting 5 3 Field module Field device 46 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 176: ...Wiring 6 5 Field control unit Field device 64 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 178: ...Operation 7 1 Operation Field device 66 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 190: ...Commissioning 8 4 Commissioning the SIPROCESS GA700 Field device 78 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 222: ...Appendix A 6 Returned delivery Field device 110 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 224: ...Index Field device 112 Operating Instructions 05 2015 A5E35640463 02 ...