Table 10-11 Temperature of measuring chamber and gas path
Temperature of measuring chamber and gas path
Gas path
5 K above ambient temperature
Standard version
Approx. 72 °C
Table 10-12 Time response
Time response
Warming-up time (at the analog output)
< 2 h
Dead time T
10
Analyzer module in the field device
(including flame arrestor):
< 1.6 s
Display delay T
90
with an electronic damping set‐
ting of 0 s and a sample gas flow of 1 Nl/min
.
Analyzer module in the
field device (including
flame arrestor)
:
≤ 2.4 s
Table 10-13 Measuring response
Measuring response
Output signal variation with static damping con‐
stant of 0 s and dynamic noise suppression of 5% /
10 s
≤ ±0.5 % of smallest measuring range
(noise bandwidth corresponds to 1 % = 6σ value
or 0.333 % = 2σ value),
with vibration compensation activated: < 1.5 times
the value
Detection limit
≤ 1% of smallest measuring range according to
module nameplate (with vibration compensation
activated:
< 1.5 times the value)
Measured value drift
At the zero point
≤ ± 0.5% of smallest measuring range / month or
≤ ± 50 vpm O
2
/ month; whichever is greater
For span gas
≤ ± 0.5% of current measured value / month or
≤ ± 50 vpm O
2
/ month; whichever is greater
Repeatability
At the zero point
≤ ± 0.5% of smallest measuring range or
≤ ± 50 vpm O
2
; whichever is greater
For span gas
≤ ± 0.5% of current measured value or
≤ ± 50 vpm O
2
; whichever is greater
Linearity deviation with ambient air
≤ 0.1% of full-scale value of characteristic
1)
1)
Untreated ambient air contains less than 20.95% O
2
(literature value) since existing humidity
decreases the oxygen content.
Table 10-14 Influence variables
Influence variables
Ambient temperature
Deviation at zero point
≤ 0.5% of smallest measuring value / 10 K or ≤ 50
vpm O
2
/ 10 K, whichever is larger
Deviation from the span
gas
≤ 0.5% of current measured value / 10 K or ≤ 50
vpm O
2
/ 10 K, whichever is larger
Technical specifications
10.3 OXYMAT 7 technical specifications
Field device
Operating Instructions, 05/2015, A5E35640463-02
91
Содержание SIPROCESS GA700
Страница 6: ...Inhaltsverzeichnis Feldgerät 6 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 22: ...Allgemeine Sicherheitshinweise 2 9 Automatisches Reset Feldgerät 22 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 64: ...Anschließen 6 5 Feldbediengerät Feldgerät 64 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 66: ...Bedienen 7 1 Bedienen Feldgerät 66 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 94: ...Technische Daten 10 3 Technische Daten OXYMAT 7 Feldgerät 94 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 112: ...Index Feldgerät 112 Betriebsanleitung 05 2015 A5E35640463 02 ...
Страница 118: ...Table of contents Field device 6 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 134: ...General safety instructions 2 9 Automatic reset Field device 22 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 158: ...Installing mounting 5 3 Field module Field device 46 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 176: ...Wiring 6 5 Field control unit Field device 64 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 178: ...Operation 7 1 Operation Field device 66 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 190: ...Commissioning 8 4 Commissioning the SIPROCESS GA700 Field device 78 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 222: ...Appendix A 6 Returned delivery Field device 110 Operating Instructions 05 2015 A5E35640463 02 ...
Страница 224: ...Index Field device 112 Operating Instructions 05 2015 A5E35640463 02 ...