Service and maintenance
12.3 Module replacement
SIMATIC RF650R/RF680R/RF685R
Configuration Manual, 03/2018, C79000-G8976-C386-06
299
XML/
LED
(hex)
OPC UA
status
AutoId
status
AutoId
text
Error description
0xE7 good
10
READ_
ERROR
Read access to the transponder has failed.
0xE8 good
14
WRITE_
ERROR
Write access to the transponder has failed.
0xE9 good
14
WRITE_
ERROR
Writing the EPC-ID on the transponder has failed.
0xEA good
1
MISC_ERROR_
TOTAL
Enabling write protection on the transponder has failed.
0xEB good
1
MISC_ERROR_
TOTAL
The "Kill" command failed.
--
good
17
DEVICE_NOT_
READY
The specified read point is inactive because it was not
assigned to any antenna.
Reading out error messages using the WBM
All the diagnostics messages of the reader are entered in the "Log" if a check mark was set
for "ERRORS" in the WBM configuration in "Settings - General". The log helps SIEMENS
specialists to analyze errors. You will find further information on the "Log" in the section "The
"Diagnostics - Log" menu item (Page 95)".
12.3
Module replacement
Before replacing a module
NOTICE
Backing up the configuration
Prior to replacing the module, make sure that you back up the configuration stored on the
reader so that you can transfer this to the newly connected reader following module
replacement.