Servo Ventilator 300/300A
Description of functions
E382 E380E 061 01 03 01
Siemens-Elema AB
29
2
2.8 Opto switch
Time multiplexed driving stage and sensor
receiving stage for all the infra-red sensor
touch pads on the front panel. The function
is controlled by the block 2.1
PANEL
CPU
including an automatic adaption of the touch
pad sensitivity to the ambient light level.
This block gives the data to the block 2.1
PANEL
CPU
.
2.9 Monitor display
This block contains all the red alpha-
numerical four character light emitting diode
displays as well as eight yellow and eight
red alarm indicating light emitting diodes
below the Alarms and messages display on
the front panel. This block is driven directly
from the main block 4
MONITORING
.
2A Automode panel
interface
The board PC 1745
AUTOMODE
PANEL
INTERFACE
contains electronics for handling of the
Automode switch.
The status of the switch is given to the
blocks 2.1
PANEL
CPU
, 5
MONITORING
and
6
REFERENCE
&
TIMING
. The LEDs are driven
from the main block 4
MONITORING
.
Содержание Servo Ventilator 300
Страница 4: ...4 Siemens Elema AB E382 E380E 061 01 03 01 Servo Ventilator 300 300A Notes Notes ...
Страница 6: ...6 Siemens Elema AB E382 E380E 061 01 03 01 Servo Ventilator 300 300A Notes Notes ...
Страница 18: ...20 Siemens Elema AB E382 E380E 061 01 03 01 Introduction Servo Ventilator 300 300A 1 Notes ...
Страница 22: ...20 Siemens Elema AB E382 E380E 061 01 03 01 Introduction Servo Ventilator 300 300A 1 Notes ...
Страница 114: ...5 112 Siemens Elema AB E382 E380E 061 01 03 01 Troubleshooting Servo Ventilator 300 300A Notes ...
Страница 130: ......