3
Servo Ventilator 300/300A
Disassembling and assembling
E382 E380E 061 01 03 01
Siemens-Elema AB
99
Cover plate
• Remove the inspiratory valves for AIR and
O
2
as previously described.
• Remove the pneumatic section casing as
previously described.
• Disconnect the flow transducer cable (1).
• Disconnect the O
2
cell cable (2).
• Remove the three screws (3).
• Carefully lift off the cover plate (4).
Expiratory valve
• Remove the cover plate as previously
described.
• Disconnect the two expiratory valve
connectors (1 and 2).
• Remove the three screws (3).
• Lift off the expiratory valve (4).
2
1
4
3
300-A44X
300-A43X
4
3
1
2