74
Siemens-Elema AB
E382 E380E 061 01 03 01
Description of functions
Servo Ventilator 300/300A
2
16 Expiratory valve
The function of this main block is to
squeeze the 14.5
EXPIRATORY
VALVE
TUBE
in a
regulated way.
16.1 Current control
Consists of the board PC 1585
CURRENT
CONTROL
.
Controller block regulating the current going
to the block 16.3
EXPIRATORY
SOLENOID
WITH
POSITION
SENSOR
according to the desired input
signal
EXP
.
VALVE
CURRENT
REF
from the block
15.2
POSITION
PD
CONTROL
. The actuating
output signal is a pulse train with constant
frequency of 30 kHz and pulse width
modulation used in the block 16.2
CURRENT
POWER
.
16.2 Current power
Consists of the board PC 1586
CURRENT
POWER
.
Driving stage for the energy going to the
block 16.3
EXPIRATORY
SOLENOID
WITH
POSITION
SENSOR
. The result of the output signal is a
certain opening of the
EXPIRATORY
VALVE
.
16.3 Expiratory solenoid with
position sensor
The
EXPIRATORY
SOLENOID
controls the opening
through the 14
EXPIRATORY
CHANNEL
by
squeezing the 14.5
EXPIRATORY
VALVE
TUBE
to
desired position.
The solenoid receives power from the block
16.2
CURRENT
POWER
. The power supply to the
solenoid is regulated so that the remaining
pressure in the patient system towards the
end of the expiration time is kept on the
PEEP level according to front panel setting.
An optical sensor on the board PC 1594
POSITION
SENSOR
E
mounted at the bottom of
the expiratory solenoid detects its position
and the resulting
EXP
.
VALVE
POSITION
signal is
used in the block 15.2
POSITION
PD
CONTROL
.
Offset zeroing for the
EXP
.
VALVE
POSITION
signal is controlled from the 6.1
MODE
CPU
via
the block 15.4
VALVE
ZEROING
.
17 Optional interface
Not used in this version.
Содержание Servo Ventilator 300
Страница 4: ...4 Siemens Elema AB E382 E380E 061 01 03 01 Servo Ventilator 300 300A Notes Notes ...
Страница 6: ...6 Siemens Elema AB E382 E380E 061 01 03 01 Servo Ventilator 300 300A Notes Notes ...
Страница 18: ...20 Siemens Elema AB E382 E380E 061 01 03 01 Introduction Servo Ventilator 300 300A 1 Notes ...
Страница 22: ...20 Siemens Elema AB E382 E380E 061 01 03 01 Introduction Servo Ventilator 300 300A 1 Notes ...
Страница 114: ...5 112 Siemens Elema AB E382 E380E 061 01 03 01 Troubleshooting Servo Ventilator 300 300A Notes ...
Страница 130: ......